5/8
Language 5.10
Entry code 5.11
Control inputs B1 and B2 5.9
NOTE! Check whether under Fct. 3.07 “HARDWARE” the output terminal “B1” and/or “B2” is
defined as control input B1 and/or B2, refer also to Sect. 2.1 and Sect. 5.16.
Fct. 1.06 and 1.07 CONTROL
B1/B2
Press key
→
twice.
Select function of control inputs,
press key
↑
.
• OFF (switched off, no function)
• OUTP. HOLD (hold value of outputs)
Functions also act on
• OUTP. ZERO (set outputs to “min. values”)
display and totalizer
• TOTAL. RESET (reset totalizers)
• ERROR. RESET (delete/acknowledge error messages)
• EXT. RANGE (external range change for automatic range change, see also Sect. 5.19,
settin
g range 5 – 80 PERCENT = ratio of low to high range 1:20 to 1:1.25,
value must be greater than that of Fct. 1.03 L.F.CUTOFF)
Press
↵
key to transfer to number setting, 1st digit (cursor) flashes.
Change flashing digit with key
↑
,
press key → to shift cursor 1 place to the right.
Press
↵
key to return to Fct. 1.06 or 1.07 CONTROL B1 or B2.
Please refer to Sect. 3.2 ”factory settings“.
Connection diagram, see Sect. 2.6.
Fct. 3.01 LANGUAGE
Press
→
key.
Select language for texts in display
•D (German)
• GB/USA (English)
•F (French)
• others on request
Select using
↑
key.
Press
↵
key to return to Fct. 3.01 LANGUAGE.
Fct. 3.04 ENTRY CODE
Press
→
key.
Choice
•NO (no code, enter setting mode with → key)
• YES (enter setting mode with → key and Code 1: → → → ↵ ↵ ↵ ↑ ↑ ↑)
Select using
↑
key.
Press
↵
key to return to Fct. 3.04 ENTRY CODE.
}