4/8
Fct. Text Description and settings
2.00 TEST Test menu
2.01 TEST Q Test measuring range Q
Precautionary query
SURE NO
Press
key to return to Fct. 2.01 “TEST Q”.
SURE YES
Press
key, then use
key to
select value:
-110 / -100 / -50 / -10 / 0 / +10 / +50 / +100 / +110 PCT.
of set full-scale range Q
100%
.
Displayed value present at outputs I and P.
Press
key to return to Fct. 2.01 “TEST Q”.
2.02 HARDW. INFO Hardware information and error status
Before consulting factory, please note down all 6 codes.
MODUL ADC X . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press
key to transfer to
“MODUL IO“.
MODUL IO X . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press
key to transfer to
“MODUL DISP.“.
MODUL DISP. X . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press
key to return to Fct. 2.02
“HARDW. INFO“.
3.00 INSTALL. Installation menu
3.01 LANGUAGE Select language for display texts
GB / USA (English) F (French)
D (German) others on request
Press
key to return to Fct. 3.01
“LANGUAGE”.
3.02 FLOWMETER Set data for primary head
DIAMETER Select size from meter size table
DN 10 - 1000 mm equivalent to 3/8 - 40 inch
Select with
key.
Press
key to transfer to subfunction “FULL SCALE”.
FULL SCALE Full-scale range for flow Q
100%
To set, refer to Fct. 1.01 “FULL SCALE” above.
Press
key to transfer to subfunction “GK VALUE”.
VALUE P Pulse value (Fct. 1.06
“VALUE P”
) has been changed.
With the “old” pulse values the output frequency (F)
would have been exceeded or not reached.
P
min
= F
min
/ Q
100%
P
max
= F
max
/ Q
100%
Check new values!
GK VALUE Set primary constant GK
see primary head nameplate.
Ran
ge: 1.0000 - 9.9999
Press
key to transfer to subfunction
“FIELD. FREQ.”.
FIELD FREQ. Magnetic field frequency
Values: 1/2, 1/6, 1/18 and 1/36
of power frequency, see nameplate.
Press
key to transfer to subfunction “FLOW DIR.”;
DC units only: to transfer to subfunction
“LINE FREQ.”.
LINE FREQ. Normal line frequency in your country
Please note: This function is only provided for units
with DC power supply to suppress line-frequency interference.
Values:
50 Hz
and
60 Hz
Press
key to transfer to subfunction “FLOW DIR.”.
FLOW DIR. Define flow direction (in F/R mode: forward flow).
Set according to direction of arrow on primary head:
+ DIR. – DIR.
Select using
key.
Press
key to return to Fct. 3.02 “FLOWMETER”.